Comparative Analysis of Absolute Dosimetry of a Varian 2300iX LINAC for Photon and Electron Beams Using IAEA TRS-398 Rev. 1

Authors

  • Sovi Anggraini Armita Dewi Universitas Pembangunan Nasional "Veteran" Jawa Timur
  • Primasari Cahya Wardhani Universitas Pembangunan Nasional "Veteran" Jawa Timur
  • Arie Chintya Martania Balai Pengamanan Alat dan Fasilitas Kesehatan (BPAFK) Surabaya

Keywords:

Absolute Dosimetry, Varian 2300iX Linac, IAEA TRS-398 Rev. 1, Radiotherapy, Quality Assurance

Abstract

The accuracy of radiation dose output is one of the factors that determines the success of radiation therapy and patient safety. Therefore, an absolute dosimetry process based on international standards is required to ensure that the dose output from the Linear Accelerator (LINAC) is consistent with the planned dose. This study aimed to analyze absolute dosimetry results on a Varian 2300iX LINAC using the IAEA TRS-398 Rev.1 protocol for 6 MV photon and 6 MeV electron beams. Measurements were performed using a water phantom, a PTW 30013 Farmer Ionization Chamber for the photon beam, a PTW 34045 Advanced Markus Chamber for the electron beam, an electrometer, a thermometer, and a barometer under reference conditions with an SSD of 100 cm, a 10 × 10 cm² irradiation field, and 200 MU. Absorbed dose to water was calculated using the absorbed-dose-to-water approach with temperature and pressure (kTP), polarity (kpol), ion recombination (ks), and beam quality (kQ) correction factors. The results showed that before adjustment, the dose-output deviations were −0.860% for the 6 MV photon beam and −0.600% for the 6 MeV electron beam. After adjustment, the deviations changed to −0.050% for the 6 MV photon beam and 0.190% for the 6 MeV electron beam. All measurement results were within the ±2% tolerance recommended by IAEA TRS-398 Rev.1. Correction factors close to one indicated that the measurement conditions met international dosimetry requirements. The results demonstrate that the Varian 2300iX LINAC has good dosimetric performance and is suitable for radiotherapy services, provided that periodic quality assurance is maintained.

Downloads

Download data is not yet available.

References

Akter, S., Khatun, R., Uddin, F., Monika, A. N., Parvej, M., Sultana, N., Hasnat, A., Alim, S., & Ahasan, M. (2022). Dosimetry Measurement of Medical Linear Accelerator and Assessment the Authentication it with Machine Records. IMedPub Journals, 7(6), 2–6. https://doi.org/10.36648/2574-285x.7.4.19.Citation

Al-saleh, W. M., & Almatani, T. (2025). Journal of Radiation Research and Applied Sciences Modelling of the 6 MV varian clinac iX linear accelerator using EGSnrc for flattening and unflattening beams. Journal of Radiation Research and Applied Sciences, 18(1), 101241. https://doi.org/10.1016/j.jrras.2024.101241

Bugani, P. S., Pulungan, M. A., & Suryana, N. (2024). Analisis Kelayakan Pengembangan Instalasi Radioterapi Pada Rumah Sakit Bersubsidi XYZ. Jurnal AL-AZHAR INDONESIA SERI SAINS DAN TEKNOLOGI, 9(1), 25–33. https://doi.org/http://dx.doi.org/10.36722/sst.v9i1.2386

Choi, D. H., Kim, J. S., Lee, R., Ahn, S. H., & Ahn, W. S. (2023). Dosimetric verification of annual quality assurance for a linear accelerator using a transmission type detector. Scientific Reports, 1–13. https://doi.org/10.1038/s41598-023-45114-2

Farrokhkish, M., Veres, A. J., Fong de Los Santos, L. E., DeMarco, J. J., & Islam, M. K. (2024). Feasibility of using a transmission ion chamber for QA tests of medical linear accelerators. Journal of Applied Clinical Medical Physics, 25(2), e14245. https://doi.org/10.1002/acm2.14245

Hasanah, H., Qomariyah, N., Makmur, I. W. A., Subroto, R., & Wirawan, R. (2020). Analisa Kurva PDD dan Dose Profile Berkas Elektron Pesawat Linac Varian Clinac CX. Indonesian Physical Review, 3(2), 84–92. https://doi.org/https://doi.org/10.29303/i pr.v3i2.43

International Atomic Energy Agency. (2024). In Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water (TRS-398 Rev.1) (Vol. 398, Issue 398). https://doi.org/DOI 10.61092/iaea.ve7q-y94k

Kawurung, R. D. O., Maslebu, G., Trihandaru, S., & Hidyatullah, M. (2018). Analisis dan Penentuan Faktor Koreksi Dosis Serap pada Medium Solid Water Phantom Terhadap Water Phantom. Jurnal Fisika FLUX, 15(1), 31–37. http://ppjp.unlam.ac.id/journal/index.php/f/

Kurnianto, P., Maslebu, G., & Muninggar, J. (2023). Quality Assurance Plan for Linear Accelerator ( LINAC ) on Nasopharynx Cancer ( NPC ) Treatment at Radiotherapy Installation Ken Saras Hospital. Advance Sustainable Science, Engineering and Technology (ASSET), 5(2), 1–9.

Meaze, A. M. H., Purohit, S., Rahman, M. S., Sattar, A., Kabir, S. E., Patwary, M. K. A., Kali, K., & Akhand, M. J. R. (2021). Studies of Dosimetry Protocols for Accelerated Photons and Electrons Delivered from Medical Linear Accelerator. East European Journal of Physics, (1), 20-26. https://doi.org/10.26565/2312-4334-2021-1-03

Sugiarta, K., Ratini, N. N., Suyanto, H., & Syarifuddin. (2022). Analisis Dosis Keluaran Berkas Sinar-X Pesawat Linac Varian Clinac CX Berdasarkan Protokol Dosimetri IAEA TRS 398. Kappa Journal, 6(2), 366–372. http://e-journal.hamzanwadi.ac.id/index.php/kpj/index

Suharmono, B. H., Anggraini, I. Y., Hilmaniyya, H., & Astuti, S. D. (2020). Quality Assurance (QA) Dan Quality Control (QC) Pada Instrumen Radioterapi Pesawat LINAC. Jurnal Biosains Pascasarjana, 22(2), 73. https://doi.org/10.20473/jbp.v22i2.2020.73-80

Sulistiana, Dewang, S., Astuty, S. D., & Saleha. (2024). ELEKTRON PADA PESAWAT LINEAR ACCELERATOR VARIAN TRUE. Berkala Fisika, 27(2), 58–68. http://ejournal.undip.ac.id/index.php/berkala_fisika

Taneja, S., & Barbee, D. L. (2024). Implementation of an ionization chamber array for linear accelerator monthly dosimetry QA. Journal of Applied Clinical Medical Physics, 25(9), e14433. https://doi.org/10.1002/acm2.14433

Tani, K., Wakita, A., Tohyama, N., & Fujita, Y. (2024). Dosimetric impact of calibration coefficients determined using linear accelerator photon and electron beams for ionization chamber in an on-site dosimetry audit. Journal of Radiation Research, 65(5), 619–627. https://doi.org/10.1093/jrr/rrae054

Downloads

Published

2026-10-16

How to Cite

Sovi Anggraini Armita Dewi, Wardhani, P. C. ., & Martania, A. C. . (2026). Comparative Analysis of Absolute Dosimetry of a Varian 2300iX LINAC for Photon and Electron Beams Using IAEA TRS-398 Rev. 1. Faraday: Journal of Fundamental Physics, Research, and Applied Science, 2(2), 11–21. Retrieved from https://faraday.upnjatim.ac.id/index.php/faraday/article/view/75

Most read articles by the same author(s)